Burkert Type 8045 Manual de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 74
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
38
Operating and functions
8.6.6. Configuring the AO1 current output
The current output gives a 22 mA current when the device shows an operation fault, even if the current
output is disabled.
The 4-20 mA output provides an electrical current, the value of which reflects the flow rate measured by the
device.
Example of relation between the measuring range and the current output:
4
20
20 180
mA
l/min
To invert the output signal, give a lower flow rate value to the 20 mA current value than to the 4 mA current
value.
To disable the current output, set both range bounds, 4 and 20 mA, to zero. In this case the output delivers a
constant current of 4 mA.
OUtPUt 4 = 28.00AO1
DO1
Enter the flow rate associated to a 4 mA current value, in the unit
chosen in the "UNIT" parameter.
Confirm.
20 = 6.000
Enter the flow rate associated to a 20 mA current value, in the unit
chosen in the "UNIT" parameter.
Confirm.
Fig. 33: Parameterizing the current output
English
Type 8045
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 73 74

Comentarios a estos manuales

Sin comentarios